ISO9000:2008 certified precision module .The XRL Series is ideal for OEM applications requiring a competitively priced, precision X-ray tube high voltage module.
Wisman's new XRL Series of X-Ray generator modules are designed for OEM applications up to 75kV to 130KV at 75-260 watts.Its universal input, small package size and choice of three standard digital interfaces(RS-232,RS-422,USB) simplifies integrating the XRL into your X-Ray analysis system. Models are available to operate either floating filament (negative HV polarity) or ground referenced filament (positive HV polarity), X-Ray tube designs. DSP based emission control circuitry pro-vides excellent regulation of emission current, along with outstanding stability performance.
TYPICAL APPLICATIONS
X-ray tubes from COMET,Kevex, Oxford, RTW, Superior, Varian and Trufocus.
OPTIONS
DC Dc input
AC AC Filament
CPC Constant Power
5VPM 0 to 5 volt programming and Monitor Scaling
TP(x) Alternate Test Point Scaling
SPECIFICATIONS
Input:
AC85-265V±10%, 4.0A maximum.
Output:
0 to 130kV at 0 to 2 mA,limited to a maximum of 75-200watts.
Voltage Control:
Local: Internal multi-turn potentiometer to set voltage from 0 to full output voltage.
Remote: 0 to +10Vdc proportional from 0 to full output voltage. Accuracy: ±1%. ZIN: 10Mohm.
Digital control:RS-422,RS-232,USB
Emission Control:
Local: Internal potentiometer to set beam current between 0 and full output.
Remote: 0 to +10Vdc proportional from 0 to full output current.
Digital control:RS-422,RS-232,USB
Accuracy : ±1%. ZIN: 10Mohm.
DC Filament Supply: Current: 5A, adjustable Voltage: 10V
Voltage Regulation:
Load: 0.01% of output voltage no load to full load. Line: ±0.01% for ±10% change in input voltage.
Current Regulation:
Load: 0.01% of output current from 0 to rated voltage. Line: 0.01% of rated current over specified input range.
Ripple:
0.1% p-p of output voltage.
Temperature Range:
0°C to +50°C operational
Temperature Coefficient:
<25PPM per °C, voltage or current regulated.
Stability:
0.05% per 8 hours after 1/2 hour warm-up.
Voltage and Current Monitors:
0 to +10Vdc proportional from 0 to rated output.Accuracy 1%.Digital control:Accuracy 2%
Dimensions:
10.14"H x 6.63"W x 3.93”D (260mm x 170mm x 100mm).
Connectors:
I/O Connectors:15 pin mini D-type Phoenix connector for
power, filament and monitor connections.
Remote Programming:
(P/O 15pin "D” analog control interface) Permits remote
adjustment of the output voltage and current via an external
potentiometer and the internal +10V reference. By adjusting
the potentiometer from minimum to maximum, the desired output may be selected.